-40%

Agilent HPLC 1260 Infinity G1365D MWD VL - Tested and working perfectly

$ 3273.6

Availability: 100 in stock
  • Return shipping will be paid by: Buyer
  • MPN: G1365D
  • Refund will be given as: Money Back
  • Brand: Agilent
  • Restocking Fee: No
  • All returns accepted: Returns Accepted
  • Item must be returned within: 14 Days
  • Condition: Unit has been tested and passed all Agilent Lab Advisor test - It is certified to Agilent specification
  • Model: 1260 Infinity

    Description

    Product Description
    The 1260 Infinity II Multiple Wavelength Detector is designed for highest optical performance, GLP compliance, and easy maintenance. With its 120 Hz data acquisition rate, the detector is perfectly suited for fast LC applications. The long-life deuterium lamps allows highest intensity and lowest detection limits over a wavelength range of 190 – 950 nm. The use of RFID tags for all flow cells and UV-lamps provides traceable information about these assemblies. The built-in holmium oxide filter features the fast wavelength accuracy verification, while the built-in temperature controls improves the baseline stability. Additional diagnostic signals for temperature and lamp voltage monitoring are available.
    Features
    Higher sensitivity and selectivity - simultaneous detection of up to eight compound-specific wavelengths. • Low detection limits - low noise front-end electronics and the patented flow cell design delivers very low detection limits thanks to the minimization of short-term noise (< ± 7 µAU). • Up to 100 % resolution gain in fast LC - using a 120 Hz data acquisition rate. • Maximum baseline stability - electronic temperature control (ETC) reduces baseline drift under fluctuating ambient temperature and humidity conditions. • Wide linear range - for reliable, simultaneous quantification of primary compounds, by-products and impurities. • Programmable slit (1 – 16 nm) for rapid optimization of sensitivity and linearity. • Excellent data traceability - radio frequency identification (RFID) tags on cells and source lamps improve traceability of data. • Automatic wavelength verification by built-in holmium oxide filter. • Nine analytical and preparative flow cells provide you with maximum application flexibility and choice. • Extensive diagnostics, error detection and display with Instant Pilot controller and Lab Advisor software.
    Performance Specifications
    Type Specification Detection type 1024-element photodiode array
    Light source Deuterium and tungsten lamps
    Number of signals 8
    Maximum data rate 120 Hz
    Short term signal noise (ASTM) < ± 0.7·10-5 AU at 254 nm and 750 nm
    Drift < 0.9·10-3 AU/h at 254 nm
    Linear absorbance range > 2 AU (5 %) at 265 nm
    Wavelength range 190 – 950 nm
    Wavelength accuracy ± 1 nm, self-calibration with deuterium lines, verification with holmium oxide filter
    Wavelength bunching 1 – 400 nm
    Slit width 1, 2, 4 , 8, 16 nm
    Diode width < 1 nm
    Time programmable Wavelength, polarity, peak width, lamp bandwidth, auto balance, wavelength range, threshold, spectra storage mode